China Small Volume Helium Gas Leak Detector Min Rate 5E-13 Display 1E-13 To 1E-1
China Small Volume Helium Gas Leak Detector Min Rate 5E-13 Display 1E-13 To 1E-1

  1. China Small Volume Helium Gas Leak Detector Min Rate 5E-13 Display 1E-13 To 1E-1

Small Volume Helium Gas Leak Detector Min Rate 5E-13 Display 1E-13 To 1E-1

  1. MOQ: 1
  2. Price:
  3. Get Latest Price
Packaging Details Wooden box for overseas shipment
Payment Terms T/T
Delivery Time 6weeks
Display 1E-13 to 1E-1 Pa•m3/s
Smallest Detectable 5E-13 Pa•m3/s
Maximum Inlet Pressure 1800Pa
Multiful interface Rs232/485, Digital I/O
Brand Name KYKY
Model Number ZQJ-3000
Place of Origin China
Certification CE

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Packaging Details Wooden box for overseas shipment Payment Terms T/T
Delivery Time 6weeks Display 1E-13 to 1E-1 Pa•m3/s
Smallest Detectable 5E-13 Pa•m3/s Maximum Inlet Pressure 1800Pa
Multiful interface Rs232/485, Digital I/O Brand Name KYKY
Model Number ZQJ-3000 Place of Origin China
Certification CE
High Light helium mass spectrometerhelium vacuum leak detector

Helium Leak Detector , ZQJ-3000 , Min rate 5*1E-13, Display 1E-13 to 1E-1

With nearly 50 years of vacuum leak detection technology experience, KYKY is the largest R&D and production base of HLD and has the independent intellectual property rights for vacuum leak detection system. The leak detectors and leak detection systems developed by KYKY provide advanced and effective leak detection solutions to aerospace, power electronics, air conditioning refrigeration, chemical metallurgy, medical equipment, semiconductor production and many other applications to meet the demands in various fields.

 

The ZQJ-3000 Helium Leak Detectors are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.

 

Advanteges:

1. Easy operation--Small volume, light weight, compact structure, simple operation panel

2. Multiful interface--Rs232/485, Digital I/O, remote controller interface, External gauge interface

3. Powerful functions--Various test mode, capability of detecting H2. 3He, amd 4He, multiple menu setting

4. Reliable performance--High sensitivity, wide measurement range, high inlet pressure, fast response time

5. Reliable quality--Prolong the service life, resistance of yttrium oxide iridium filament

 

Specifications:

Type ZQJ-3000
Smallest Detectable Leak Rate(Pa•m3/s) 5×10-13
Leak Rate Display(Pa•m3/s) 10-13~10-1
Maximum Inlet Pressure(Pa) 1800
Response Time(s) ≤0.3
Run-up Time(min) <3
Power 230 VAC±10%/50 Hz
120V±10%/60 Hz,10A
Working Temperature and Relative humidity Working Temperature 10~35℃,realtive humidity ≤80%
L*W*H(mm) 550×460×304
Weight(kg) 44

 

Methods:

Vacuum Method

In the vacuum method test gas is blown against the wall of the evacuated sample from the atmosphere side. It enters the sample at leaks and is fed to the leak detector.

The sample must be vacuum pressure-proof.

The sensitivity stages GROSS ---FINE---ULTRA are run through.

The detection limit is lower than in the sniffing method. The helium concentration at the leak must be known in order to quantify the leak. The state of equilibrium must be waited for.

 

Sniffing Method

In the sniffing method the test gas escaping from leaks in the sample into the atmosphere is detected.

The sample must withstand the applied test pressure.

In operation with the sniffing probe a constant gas flow is sucked in from the atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display of approx. 1*10-6 mbar l/s which can be eliminated by the ZERO function.20 3 Description Operating Instructions, ikna88en1-01, 1605

 

To detect a leak, the sniffing probe is applied to the points of the sample under helium overpressure which are suspected of leaking. An increased leak rate value indicates an increased concentration of helium and therefore a leak. The higher the pressure and the helium concentration in the sample, the smaller the leaks which can be detected.

The sensitivity stages GROSS---FINE are run through.

The detection sensitivity and the quantifiability of the leak rate are less favourable than in the vacuum pressure leak detection.

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer

  • Year Established:

    1958

  • Employee Number:

    >500

  • Ecer Certification:

    Site Member

KYKY TECHNOLOGY CO., LTD., was founded in 1958, the pioneer of vacuum technology and electron optics in China. In the past 60 years, KYKY has been dedicated to providing comprehensive vacuum solutions to the customers all over the world.   KYKY invented the first set of Turbo Molecular Pu... KYKY TECHNOLOGY CO., LTD., was founded in 1958, the pioneer of vacuum technology and electron optics in China. In the past 60 years, KYKY has been dedicated to providing comprehensive vacuum solutions to the customers all over the world.   KYKY invented the first set of Turbo Molecular Pu...

+ Read More

Get in touch with us

  • Reach Us
  • KYKY TECHNOLOGY CO., LTD.
  • No. 13 Zhongguancun Bei er tiao, Haidian District, Beijing,100190
  • https://www.kykytech.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement