Payment Terms | L/C, T/T, Western Union |
Supply Ability | 100 sets per month |
Delivery Time | 30 days |
Packaging Details | ASTM D6184 Oil Separation Kit packed in hardboard |
Test standard | ASTM D6184 |
Sieve | 60 mesh |
Operation | Manual |
Test object | Lubricating Grease |
Weight | 1kg |
Warranty | 1 Year |
Brand Name | AI |
Model Number | AIC-6184 |
Certification | ISO |
Place of Origin | China |
View Detail Information
Explore similar products
ASTM D127 Drop Melting Point Apparatus for Petroleum Wax and Grease
ASTM D1478 Low-Temperature Torque Test Equipment with Standard 204 Bearing
Anti-rust Grease Humidity Cabinet for Rust Protection by Metal Preservatives
ASTM D4290 Leakage Tendencies Tester For Automotive Wheel Bearing Grease
Product Specification
Payment Terms | L/C, T/T, Western Union | Supply Ability | 100 sets per month |
Delivery Time | 30 days | Packaging Details | ASTM D6184 Oil Separation Kit packed in hardboard |
Test standard | ASTM D6184 | Sieve | 60 mesh |
Operation | Manual | Test object | Lubricating Grease |
Weight | 1kg | Warranty | 1 Year |
Brand Name | AI | Model Number | AIC-6184 |
Certification | ISO | Place of Origin | China |
High Light | ASTM D6184 ,Oil Separation Kit ,Conical Sieve D6184 |
ASTM D6184 Oil Separation Kit is suitable for measuring the tendency of lubricating grease to separate oil at higher temperatures.
Place approximately 10 grams of the sample in a metal wire mesh and let it stand still at 100 ℃+1 ℃. After 30 hours, measure the mass percentage of oil flowing through the sieve.
Apparatus and materials
1. Nickel wire steel mesh or stainless steel mesh: 60 mesh, with a metal wire lifting beam above the steel mesh.
2. Beaker: tall, mouthless, 200 ml.
3. Lid: Tightly fitting with the cup, there should be a hook at the center of the lid bottom.
4. Thermostatic box: the temperature can be controlled to 1 ℃.
Experimental procedure
1. Preheat the constant temperature box to 100 ℃+1 ℃. (or at a specified temperature)
2. Weigh a clean beaker (without steel mesh and lid) to an accuracy of 0.01 grams.
3. With minimal stirring, carefully place about 10 grams (accurate to 0.01 grams) of the sample into the steel mesh using a scraper (the steel tip should be filled with the sample), avoiding the formation of bubbles and voids, making the top of the sample smooth and convex to prevent oil accumulation. Use your fingers to wipe off the sample squeezed out from the mesh, weigh the steel mesh that has been loaded with the sample, accurate to 0.01 grams..
4. Place the assembled steel mesh oil separator into a constant temperature box and maintain it at 100 ℃+1 ℃ for 30 hours.
5. After reaching the specified time, remove the steel mesh oil separator from the constant temperature box and cool it to room temperature.
6. Take out the steel mesh from the cup and let the oil on the tip of the mesh drip into the beaker along the cup wall.
7. Weigh the mass of oil in the beaker to the nearest 0.01 grams.
Company Details
Business Type:
Manufacturer,Trading Company
Year Established:
2018
Total Annual:
1,000,000-5,000,000
Employee Number:
150~200
Ecer Certification:
Verified Supplier
Advanced Instruments Co.,Limited. is a professional supplier which is committed to offer best testing equipment from China for our clients,and our main products are fire rating testing equipment,petroleum products testing equipment,electrical testing equipment geophysical equipment and general labor... Advanced Instruments Co.,Limited. is a professional supplier which is committed to offer best testing equipment from China for our clients,and our main products are fire rating testing equipment,petroleum products testing equipment,electrical testing equipment geophysical equipment and general labor...
Get in touch with us