High Accuracy Portable Hiv Test Kit Plasma Specimen Clean Sterile Gauze Pad

  1. MOQ: 5000pcs
  2. Price: USD0.25-USD1
  3. Get Latest Price
Payment Terms T/T, Western Union
Supply Ability 2000000pcs/month
Delivery Time 20-30days
Packaging Details 1pc/pouch, 25pcs/box
Method Gold Colloidal
Format Home kit
Specimen Whole Blood/Serum/Plasma
Testing time 5-15 minutes
Shelf Life 24 Months
Application HIV
Brand Name New Life
Certification ISO13485,
Place of Origin China

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Payment Terms T/T, Western Union Supply Ability 2000000pcs/month
Delivery Time 20-30days Packaging Details 1pc/pouch, 25pcs/box
Method Gold Colloidal Format Home kit
Specimen Whole Blood/Serum/Plasma Testing time 5-15 minutes
Shelf Life 24 Months Application HIV
Brand Name New Life Certification ISO13485,
Place of Origin China
High Light home medical test kitshome health test kits

HIV Self Test kit , Certified, Whole Blood/Serum/Plasma Specimen, Home test use for HIV antibody

 

 

Accessories:

 

 

Test Cassettes Antiseptic wipe Sterile lancet
Pippette Sterile gauze pad Package Insert
Buffer    

 

 

 

INTENDED USE:

 

 

The HIV self Test kit (Whole Blood/Serum/Plasma) is a rapid chromatographic immunoassay for the qualitative detection of antibodies to Human Immunodeficiency Virus(HIV) type 1 and type 2 in whole blood, serum or plasma to aid in the diagnosis of HIV infection. It is easily for personally use at home.

 

 

SUMMARY AND EXPLANATION OF THE TEST

 

HIV is the etiologic agent of Acquired Immune Deficiency Syndrome (AIDS). The virion is surrounded by a lipid envelope that is derived from host cell membrane. Several viral glycoproteins are on the envelope. Each virus contains two copies of positive-sense genomic RNAs. HIV 1 has been isolated from patients with AIDS and AIDS-related complex, and from healthy people with high potential risk for developing AIDS.1 HIV 2 has been isolated from West African AIDS patients and from seropositive asymptomatic individuals.2 Both HIV 1 and HIV 2 elicit immune response.3 Detection of HIV antibodies in serum, plasma is the most efficient and common way to determine whether an individual has been exposed to HIV and to screen blood and blood products for HIV.4 Despite the differences in their biological characteristics, serological activities and genome sequences, HIV 1 and HIV 2 show strong antigenic cross-reactivity.5,6 Most HIV 2 positive sera can be identified by using HIV 1 based serological tests. The HIV 1.2 Rapid Test Cassette(Whole Blood/Serum/Plasma) is a rapid test to qualitatively detect the presence of antibody to HIV 1 and/or HIV 2 in whole blood, serum or plasma specimen. The test utilizes latex conjugate and multiple recombinant HIV proteins to selectively detect antibodies to the HIV 1.2 in whole blood, serum or plasma.

 

 

Test Principle

 

The HIV 1.2 Rapid Test Cassette (Whole Blood/Serum/Plasma) is a qualitative, membrane based immunoassay for the detection of antibodies to HIV 1.2 in whole blood, serum or plasma. The membrane is pre-coated with recombinant HIV antigens. During testing, the whole blood, serum or plasma specimen reacts with HIV antigen coated particles in the test Dipstick. The mixture then migrates upward on the membrane chromatographically by capillary action and reacts with recombinant HIV antigen on the membrane in the test line region. If the specimen contains antibodies to HIV 1 and/or HIV 2, a colored line will appear in the test line region, indicating a positive result. If the specimen does not contain HIV 1 and/or HIV 2 antibodies, a colored line will not appear in the test line region, indicating a negative result. To serve as a procedural control, a colored line will always appear in the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

 

TEST PROCEDURE


Allow the test, specimen, buffer and/or controls to reach room temperature (15‐30) prior to testing.

 

 

1.  Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible.

 

2. Place the cassette on a clean and level surface.

 

For Venipuncture Whole Blood specimen: Hold the dropper vertically and transfer 2 drops of whole blood (approximately 50 mL) to the specimen area, then add 2 drops of buffer (approximately 80mL), and start the timer. See illustration below.

 

For Fingerstick Whole Blood specimen:

 

 

INTERPRETATION OF RESULTS


Negative: The presence of only one line in the control region indicates a negative result (Figure 1).

 

HIV-1 Positive: The control line and HIV-1 line (T1) are visible in the result window. The test is positive for HIV-1.


HIV-2 Positive: The control line and HIV-2 line (T2) are visible in the result window. The test is positive for HIV-2.


HIV-1 and HIV-2 Positive: The control line, HIV-1 (T1) and HIV-2 (T2) lines are visible in the result window. The test is positive for HIV-1and HIV-2.

 

Regarding the positive results for both HIV-1 and HIV-2 in one patient, it is possible for reasons as follows:

 


1.  There is the homology in the amino acid sequence of HIV type-1 and type-2. So, it is possible that the test
results show the positive results for HIV-1 and HIV-2 in one patient, simultaneously.
2.  Provisionally, you can conclude virus type according to the line density. If the line density of type-1 is darker
than that of type-2 in the result window, you can read as HIV-1 positive. If the line density of type-2 is darker
than that of type-1 in the result window, you can read as HIV-2 positive. If you want to determine virus type or co-infection exactly, you should perform the confirmatory assay (e.g Western blot etc.).


Invalid: If the pink color line in C region is not visible, the result is considered invalid (Figure 3) regardless of the presence or absence of the test line(s).

 

 

Interfering Substance

 

 

Analytes were spiked into negative plasma and serum pools (ELISA confirmed) and low positive plasma and serum specimens (ELISA confirmed) at the concentrations listed. The specimens were tested in triplicate with visual interpretations occurring at 10 and 20 minutes after specimen application. Results are presented in table below.

 

Table: Interfering Substance

 

    HIV14050004‐T   HIV14050005‐T   HIV14050006‐T  
  Concentratio                        
Analytes Plasma Serum Plasma Serum Plasma Serum
n                              
  Neg P‐ Neg   S‐ Neg P‐ Neg   S‐ Neg P‐ Neg   S‐
         
      0 0   0 0   0
    . 03 .   . .   . .  
      5 3   5 3   5
                       
Ascorbic acid 20mg/ml ‐* +*   + +   + +   +
*      
                               
                                 
Hemoglobin 1,100mg/dl +   + +   + +   +
                                 
Gentistic acid 20mg/dl +   + +   + +   +
                                 
Oxalic acid 600mg/dl +   + +   + +   +
                                 
Bilirubin 1000mg/dl +   + +   + +   +
                                 
Acetaminophe 20mg/dl +   + +   + +   +
n      
                               
                                 
Acetylsalicylic 20mg/dl +   + +   + +   +
Acid      
                               
                                 
Creatine 200mg/dl +   + +   + +   +
                                 
Albumin 2000mg/dl +   + +   + +   +
                                 
Caffeine 20mg/dl +   + +   + +   +
                               
Note: “*” mean negative result “**” mean positive result                  

 

 

Conclusion:

 

 

As per the results tabulated above, no substances showed any interference with the expected test results within the specified result interpretation time of 10 to 20 minutes.

 

 

ORIENT NEW LIFE MEDICAL CO., LTD.
Contact: Jerry Meng
Email: Jerry @ newlifebiotest .com
Tel. +86 18657312116
SKYPE enetjerry

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer,Exporter

  • Year Established:

    2009

  • Total Annual:

    10,000,000-15,000,000

  • Employee Number:

    200~300

  • Ecer Certification:

    Active Member

Established in 2009, Orient New Life Medical Co., Ltd., is specialized in high-tech biological diagnostic test development, research and production, as well as high-tech medical device development, research and production. New Life has built up two production base, which produce biological diagnosti... Established in 2009, Orient New Life Medical Co., Ltd., is specialized in high-tech biological diagnostic test development, research and production, as well as high-tech medical device development, research and production. New Life has built up two production base, which produce biological diagnosti...

+ Read More

Get in touch with us

  • Reach Us
  • Orient New Life Medical Co.,Ltd.
  • 12F,Guting Building,Hefeng Creative Square,No.495 Jiangdong North Road,Ningbo,Zhejiang, China
  • https://www.medicaldiagnostictestkits.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement