Payment Terms | L/C, D/A, D/P, T/T |
Packaging Details | Carton |
Maximum centrifugal force (×g) | 2600 ×g |
Maximum speed(r/min) | 4000 r/min |
noise | ≤60dB |
Early warning function | Overspeed、overtemperature、imbalance、etc |
Weight | 18000 g |
Dimensions (length×width×height) | 38×26×31 cm |
Brand Name | HAIYING |
Model Number | TD4 |
Place of Origin | China |
View Detail Information
Explore similar products
TDZ5 4×100mL Low Speed Centrifuge Centrifugal Precipitation
Intelligent Control Lab Centrifuge Machine 6500 Rpm
TD4 8×20mL Tabletop Low Speed Centrifuge Rpm 4000
Three Layer Steel Protective Sleeve Speed Centrifuge TDZ4
Product Specification
Payment Terms | L/C, D/A, D/P, T/T | Packaging Details | Carton |
Maximum centrifugal force (×g) | 2600 ×g | Maximum speed(r/min) | 4000 r/min |
noise | ≤60dB | Early warning function | Overspeed、overtemperature、imbalance、etc |
Weight | 18000 g | Dimensions (length×width×height) | 38×26×31 cm |
Brand Name | HAIYING | Model Number | TD4 |
Place of Origin | China | ||
High Light | 2600 Xg Medical Laboratory Centrifuge ,Sturdy Safety Medical Laboratory Centrifuge ,2600 Xg blood centrifuge machine |
Host technical parameters | |
product name | Sturdy Durable Safety Maximum Centrifugal Force 2600 Xg TD4 Low Speed Centrifuge |
Maximum capacity | 8×20 ml |
power | 120W |
power supply | AC220V 50Hz |
Weight (with rotor) | 18 Kg |
Dimensions (length×width×height) | 380×260×310 mm |
TD4 Low Speed Centrifuge Common failure analysis
After the power is turned on, the control panel displays normal, and the cooling fan works normally. When the start button is pressed, the intermediate relay works and the centrifuge does not work. According to the analysis of the fault phenomenon, the power supply part and the control part should be normal, and the fault should be at the line from the control board to the motor, the carbon brush and the motor. Turn on the centrifuge, use a multimeter to measure the continuity of the line from the control board to the carbon brush, and then measure that the carbon brush does not conduct when it reaches the motor rotor. Carefully observe the poor contact between the carbon brush and the motor rotor. Remove the carbon brush and find that it has been used for too long. Caused by too short. Replace with a new carbon brush, troubleshooting. This kind of failure is a common failure of the centrifuge, and the carbon brushes should be replaced regularly to avoid the above failures.
Company Details
Business Type:
Distributor/Wholesaler,Trading Company
Year Established:
2016
Total Annual:
1800000-2000000
Employee Number:
80~90
Ecer Certification:
Active Member
The company uses the business philosophy that is fully integrated with the market economy and efficient management and operation mechanism, takes the market as the guide, and builds a "customer-centric" marketing system based on customer needs. Ensure to provide customers with high-quality... The company uses the business philosophy that is fully integrated with the market economy and efficient management and operation mechanism, takes the market as the guide, and builds a "customer-centric" marketing system based on customer needs. Ensure to provide customers with high-quality...
Get in touch with us
Leave a Message, we will call you back quickly!